Thermocouple,Input,Field,Rangeable,DIN Rail,Isolated,Transmitter,Model DM4130,Wilkerson Instrument

Thermocouple Input Field Rangeable DIN Rail Isolated Transmitter Model DM4130


TERMINAL CONNECTIONS
Power AC L1 or DC +
Power AC L2 or DC -
T/C Input +
T/C Input -
Shield
Output +
Output -
(Terminal Labeled)
L1
L2
1
2
3
Output +
Output -
Thermocouple,Input,Field,Rangeable,DIN Rail,Isolated,Transmitter,Model DM4130,Wilkerson Instrument

Thermocouple,Input,Field,Rangeable,DIN Rail,Isolated,Transmitter,Model DM4130,Wilkerson Instrument

Features

Thermocouple,Input,Field,Rangeable,DIN Rail,Isolated,Transmitter,Model DM4130,Wilkerson Instrument Spans from 6.25% to 100% of T/C Range
Thermocouple,Input,Field,Rangeable,DIN Rail,Isolated,Transmitter,Model DM4130,Wilkerson Instrument Linearized for J, K, R, S, T, E, and N Thermocouples
Thermocouple,Input,Field,Rangeable,DIN Rail,Isolated,Transmitter,Model DM4130,Wilkerson Instrument Provides a Fully Isolated DC Output Proportional to T/C Input
Thermocouple,Input,Field,Rangeable,DIN Rail,Isolated,Transmitter,Model DM4130,Wilkerson Instrument Plug-in Output Terminal Block
Thermocouple,Input,Field,Rangeable,DIN Rail,Isolated,Transmitter,Model DM4130,Wilkerson Instrument Standard DIN rail Mount With Easy Snap-on Snap-off
Thermocouple,Input,Field,Rangeable,DIN Rail,Isolated,Transmitter,Model DM4130,Wilkerson Instrument Connections and Ranging Readily Accessible From The Front
Thermocouple,Input,Field,Rangeable,DIN Rail,Isolated,Transmitter,Model DM4130,Wilkerson Instrument User-Settable Input and Output Ranges
Thermocouple,Input,Field,Rangeable,DIN Rail,Isolated,Transmitter,Model DM4130,Wilkerson Instrument Fully Labeled Jumper Positions for Easy Range Settings
Thermocouple,Input,Field,Rangeable,DIN Rail,Isolated,Transmitter,Model DM4130,Wilkerson Instrument Choice of Power Options
Thermocouple,Input,Field,Rangeable,DIN Rail,Isolated,Transmitter,Model DM4130,Wilkerson Instrument User Write-on Label
Thermocouple,Input,Field,Rangeable,DIN Rail,Isolated,Transmitter,Model DM4130,Wilkerson Instrument Permanent Warranty
Thermocouple,Input,Field,Rangeable,DIN Rail,Isolated,Transmitter,Model DM4130,Wilkerson Instrument UL, cUL Recognized

Description

The DM4130 provides an isolated voltage or current output proportional to temperature as measured with a thermocouple. Any thermocouple may be used, and types J, K, R, S, T, E, and N can be linearized.

A microprocessor controls span and offset settings, performs cold junction compensation calculations, and performs a table look-up for linearization.

The DM4130 includes filtering and conditioning to reduce susceptibility to transients and noisy operations. It utilizes a feedback VCO to develop a pulse train with a duty cycle proportional to the input signal amplitude. This pulse train is coupled through a pulse transformer to the output circuitry, where the duty cycle data is converted to a proportional DC output level.

Connections are made via two 4-pin terminal blocks. The lower terminal block is unpluggable, allowing the product to be removed without unwiring the output and power connections.

Typical Applications

Useful for temperature monitoring, control  and data acquisition.

Mounting

The module mounts on a standard H-35 DIN rail. A spring latch holds it in place. The module is demounted by using a screw driver to release the latch. The latch release is accessible at the bottom of the module.

Specifications

INPUT
Minimum
  Span Adjustment
  (5 mV input minimum)

Maximum
  100% of T/C range

Offset Adjustment
  0% to 90% of T/C range

LINEARIZATION
±0.25°C of NIST tables
full range of J, K, R, S, T, E, N

COLD JUNCTION COMPENSATION
+\-0.25°C accuracy

IMPEDANCE
1 megohm

UPDATE PERIOD
4/sec

POWER
Wattage 2.5 W maximum 
 
AC (Standard)
  115 VAC  ±10%, 50/60 Hz
  230 VAC  ±10%, 50/60 Hz
  24 VAC   ±10%, 50/60 Hz

DC (Optional)
  24 VDC   (21-32 VDC)

APPROVALS
UL, cUL Certified

OUTPUT
Voltage
0/.25
0/1
1/5
0/5
0/10
-5/5
-10/10
Current (mA)
0/1
0/4
4/20
0/20

OUTPUT LOGIC
Normal or Reverse Acting

DRIVE CAPABILITY
Voltage 
  5 mA
Current 
  >24 V compliance

ACCURACY
±0.1% of span

OUTPUT RIPPLE
<0.1% of full scale
  (peak to peak)

OPERATING TEMPERATURE
 14 to 140°F
-10 to  60°C

TEMPERATURE STABILITY
±(.01% of span)/°C

ISOLATION
Resistance
  >500 megohms
Voltage Rating
  >1000 VAC Sinewave

COMMON MODE REJECTION
>120 dB, DC to 60 Hz